3 Spectrum Analyzer Mode
3.10 TOI Measurement
Upper and Lower Tone (set to Sense)
Zero span measurement Resolution Bandwidth
Zero span measurement Dwell Time
Harmonics (SA Mode only)
Parameters affected by Auto Couple are:
Resolution Bandwidth
Fundamental Frequency
Dwell Time
Range Table Resolution Bandwidths
Range Table Dwell Times
Meas Preset
Initiates the first sweep. Returns Meas Local variables in the current measurement
to their preset values (many of which are determined from the results of the first
sweep).
Remote Command
:CONFigure:TOI
3.10.8.2 Zero Span
Contains Meas Setup functions that enable you to control Zero-Span measurement
parameters.
Zero-Span Measurement
Zero-Span Measurement uses two additional zero-span sweeps one at each
intermodulation product to significantly increase the accuracy and the dynamic
range of the TOI measurement. Zero-span measurement cannot be used on signals
that vary rapidly in frequency.
Remote Command
[:SENSe]:TOI:ZSPan:STATe ON | OFF | 1 | 0
[:SENSe]:TOI:ZSPan:STATe?
Example
:TOI:ZSP:STAT OFF
Spectrum Analyzer Mode User's &Programmer's Reference 1831