3 Spectrum Analyzer Mode
3.9 SEM Measurement
n Data
# Item
1 Reserved for future use, returns -999.0
2 Reserved for future use, returns -999.0
3
2k + 3, k = 0
At negative offset A
4
2k + 4, k = 0
At positive offset A
---
26
2k + 4, k = 11
At positive offset L
3.9.6 Results for n = 12
When "Measurement Type" on page 1700 is Spectrum Peak reference, returns the
peak power of the signal in the ref channel
Otherwise, the value returned is -999.0
3.9.7 Results for n = 14
Returns comma-separated scalar results
For MSR, 5G NR, LTEAFDD, and LTEATDD Modes, returns outer offset results when
"Non-Contiguous Meas Region" on page 1622 is set to Outer or Outer & Inner
(5GNR, LTEAFDD, LTEATDD only), and returns inner offset results when it is set to
Inner
The results are in the following order:
k is an index for each offset: k = 0 for offset A; k = 1 for offset B; k = 2 for offset C, etc.
The number of offsets is 12 (A-L)
# Item Unit
1
10k + 1, k
= 0
Relative integrated power on the negative offset A
Returns NaN when "Measurement Type" on page 1700 is
Spectrum Peak Ref
When "Measurement Type" on page 1700 is
Total Power Ref: dBc
When PSD: dB
2
10k + 2, k
= 0
Absolute integrated power on the negative offset A
Returns NaN when "Measurement Type" on page 1700 is
Spectrum Peak Ref
When "Measurement Type" on page 1700 is
Total Power Ref: dBm
When PSD: dBm/Hz
3
10k + 3, k
Relative peak power on the negative offset A When "Measurement Type" on page 1700 is
Total Power Ref PSD: dBc
1534 Spectrum Analyzer Mode User's &Programmer's Reference