3 Spectrum Analyzer Mode
3.10 TOI Measurement
Avg/Hold Num
When Average/Hold is turned on, at each frequency a specified number of sweeps is
taken, and the data in those sweeps averaged for the purposes of establishing the
measurement.
Remote Command
[:SENSe]:TOI:AVERage:COUNt <integer>
[:SENSe]:TOI:AVERage:COUNt?
Example
:TOI:AVER:COUN 20
Preset 10
Min 1
Max 10000
Annotation In the Meas Bar
Backwards Compatibility
SCPI
[:SENSe]:TOINtercept:AVERage:COUNt
Auto Function
Remote Command
[:SENSe]:TOI:AVERage[:STATe] ON | OFF | 1 | 0
[:SENSe]:TOI:AVERage[:STATe]?
Preset
OFF
Backwards Compatibility
SCPI
[:SENSe]:TOINtercept:AVERage[:STATe]
Averaging On/Off
Turns Averaging on or off.
NOTE
In this measurement, the Average Type is preset to the Log-Pwr Avg (Video)
method. Other averaging methods are not available.
Remote Command
[:SENSe]:TOI:AVERage[:STATe] OFF | ON | 0 | 1
[:SENSe]:TOI:AVERage[:STATe]?
Example
:TOI:AVER OFF
:TOI:AVER?
Preset
ON
State Saved Yes
Range
OFF|ON
1828 Spectrum Analyzer Mode User's &Programmer's Reference