40 U9361C/F/G/M RCal Measurement Guide
Making Measurements
Measuring Magnitude Cal with Multiple States
23. Clear the Calibrate checkbox on row 1 and then select Calibrate for Row 2
and then select Calibrate Checked Rows.
24. When the calibration is complete, select Apply and close the Calibration
Configuration window.
25. Disconnect the cable from the RCal module and connect it to the signal
generator.
26. Select Meas Setup > Source > RF Output > On.
Now the trace should be properly corrected without errors.
27. View the new trace by adding a 3rd trace.
Trace > Select Trace > Trace 3
Trace Type to Max Hold
View to Active
Observe the differences between the uncorrected trace (Trace 1), the
corrected trace (Trace 2), and the corrected trace with Preselector Bypass
enabled (Trace 3).