U9361C/F/G/M RCal Measurement Guide 45
Making Measurements
Using Complex Cal to Improve EVM in a 128 QAM Signal
13. Select the Calibrate checkbox for Row 1 and then Calibrate Checked
Rows > OK.
14. When calibration is complete, select Apply.
15. Disconnect the cable from the RCal module and connect it to the signal
generator.
16. Select RF Output > On.