U9361C/F/G/M RCal Measurement Guide 73
Making Measurements
Using RCal Directly from the 89600 VSA
The results have been tabulated below. TX power is -10 dBm.
The average difference in results between the Power Sensor and the
measurement using an RCal calibration for this test case was 0.192 dBm
across the selected measurement span.
27. Before disconnecting the power sensor, select Return > Return > and set
Channel A Off.
Frequency (GHz) Power Sensor (dBm) RCal Corrected Difference (Abs)
2.5 -10.07 -10.01 0.06
5 -10.06 -9.996 0.064
8 -10.35 -9.881 0.469
11 -9.97 -10.08 0.11
14 -9.91 -10.16 0.25
17 -10.12 -10.08 0.04
20 -10.36 -10.01 0.35