U9361C/F/G/M RCal Measurement Guide 77
Making Measurements
Using RCal as a Scalar Network Analyzer
12. Select Trace (Trace 1) View/Blank to View.
13. Disconnect the RCal RF Output from the signal analyzer RF Input and add
an attenuator to the output of the RCAL. For this example, we are using a
3 dB attenuator. Connect the other end of the attenuator to the RF input of
the signal analyzer.
14. Select Trace> Trace 2 > View/Blank to Active.
15. Select Trace 2 > Trace Type to Max Hold.
16. Select Input/Output > Calibrator Control > Frequency to 4 GHz and scroll
the RPG know slowly 200 times until the Frequency reads 4.210 GHz.
Alternatively, you can use the right arrow key.
This will trace out the frequency response across all 201 points of the RCal
plus the attenuator.