76 U9361C/F/G/M RCal Measurement Guide
Making Measurements
Using RCal as a Scalar Network Analyzer
9. Set Output On.
10.Select Trace > Trace Type > Max Hold.
11. Select Input/Output > Calibrator Control > Frequency and scroll the RPG
slowly 200 times until the Frequency reads 4.210 GHz. (If you go too fast,
you will see steps in the trace.) Alternatively, you can use the right arrow
key.