72 PMC-A
6.
Turn the VOLTAGE knob clockwise until constant
current operation is achieved.
Offset adjustment is always performed under constant current
operation.
7.
Adjust Iout OFS so that the output current (a value
calculated from the external DVM reading and shunt
resistance) is 0 A.
Type I: Variable resistor D
Ammeter offset (Type I model only)
8.
Adjust I METER OFS so that the ammeter reading on
the power supply indicates 0.
Type I: Variable resistor 3
Output current at full scale
 ・Do not set the output current greater than or equal
 to 105 % of the rated current for this adjustment. If
 you do, the power supply may malfunction.
If the output current is being controlled using an external
voltage (CC-V mode) on Type II models, the output cur-
rent at full scale cannot be adjusted using the variable
resistor. Adjust the external voltage to obtain the full scale
current.
9.
Set the output current to the rated output current.
When using local control, turn the CURRENT knob clock-
wise all the way. When using analog remote control, set
the control signal to 10 V or 10 k
Ω.
10.
Adjust Iout MAX so that the output current (a value
calculated from the external DVM reading and shunt
resistance) is slightly higher than the rated current.
Both Type I and Type II: Variable resistor B
CAUTION