88
Specications
HOLD input Galvanically separated (optocoupler)
Function Switches device to HOLD mode
Switching voltage 0 ... 2 V (AC/DC) HOLD inactive
10 ... 30 V (AC/DC) HOLD active
Output 1 0/4 ... 20 mA, max. 10 V, floating
(terminals 8 / 9, galvanically connected to output 2)
Overrange * 22 mA in the case of error messages
Characteristic Linear, bilinear or logarithmic
Output filter * PT
1
filter, time constant 0 ... 120 s
Measurement error
1)
< 0.25% current value + 0.025 mA
Output 2 0/4 ... 20 mA, max. 10 V, floating
(terminals 9 / 10, galvanically connected to output 1)
Overrange * 22 mA in the case of error messages
Characteristic Linear, bilinear or logarithmic
Output filter * PT
1
filter, time constant 0 ... 120 s
Measurement error
1)
< 0.25% current value + 0.025 mA
*)
user-defined
1)
according to IEC 746 Part 1, at normal operating conditions