Specications
Relays 1 / 2 Two relay contacts, floating (terminals 14 / 15 / 16)
Contact ratings AC < 250 V / < 3 A / < 750 VA
DC < 30 V / < 3 A / < 90 W
Usage Limit value
Alarm
Wash
Limit value Function Min or Max
Setpoint As desired within range
Contact response N/C or N/O
Hysteresis User-defined
Response delay 0000 ... 9999 s
Alarm Trigger Failure or Sensoface
Contact response N/C or N/O
Wash Cycle time 0.1 ... 999.9 h
ON time 0 ... 1999 s
Contact response N/C or N/O
Real-time clock Different time and date formats selectable
Power reserve > 5 days
Display LC display, 7-segment with icons, colored backlighting
Primary display Character height approx. 22 mm, unit symbols approx. 14 mm
Secondary display Character height approx. 10 mm
Text line 14 characters, 14 segments
Sensoface 3 status indicators (friendly, neutral, sad face)
Mode indicators meas, cal, conf, diag
Further icons for configuration and messages
Alarm indication Display blinks, red backlighting
Keypad Keys: meas, info, 4 cursor keys, enter
Diagnostics
Calibration data Calibration date, zero, slope
Device self-test Automatic memory test (RAM, FLASH, EEPROM)
Display test Display of all segments
Logbook 100 events with date and time