6H4130 – September 2007 27
Section 5: Appendix A: Tuning Process Results Form
Table 1: Factory X-ray Parameters Used to Tune the CR7400 unit.
IMAGING MODE X-RAY TUBE SID EXPOSURE TIME
Intraoral 70kV, 8mA 12” 0.16s
cephalometric 70kV, 10mA 150cm 0.17s
Panoramic 70kV, 8mA 45.5cm 12s
Tuning Process Results
Machine Serial Number:
Software version:
Date:
Tested by:
Signature:
X-ray Exposure Conditions:
Parameter
Intraoral cephalometric Panoramic
Units
Factory Actual Factory Actual Factory Actual
X-ray tube voltage
and current
70; 8 70; 10 70; 8 kVp; mA
Exposure time 0.16 0.75 12 sec.
FSFD 12 in. 150 cm 45.5 cm Data C3
Screen identification Data C4
Screen format 31 x 41 Ceph Pan mm x mm