11 Menu description
11.4 Menu 2: General settings
Menu item Description
1.09Display direction
change VolumeterB
Shows the number of the flow direction changes of the flowmeterB since the last reset.
Requirement:
2.13Setting Function pulse inputs Set to Encoder .
Note:
The counter can be increased in case of a power failure even without change of direction. Recom-
mendation: Reset the counter before starting the measurement.
1.10Display reset
bypass and group error
message
Reset the group error message after eliminating the cause for the activation of the collective error
message and bypass valve.
1.11Setting Display
brightness
Adjust the brightness of the display
1.12Setting Display
contrast
Adjust the contrast of the display
1.13Setting Select
language
Selection of the language
1.30Setting Enter
password
Allows password entry, can only be reached via 2.01Setting Enable password or 2.02Setting
Change password.
1.31Help on operation Shows the brief instruction.
11.4 Menu 2: General settings
o Changing general settings in accordance with the requirements of the measuring task
o Changes only possible with password
o Ä Operation,Page22
Menu item Description
2.01 Setting Enable
Password
Factory settings:
o Password: 1000
o Password protection: Yes
Yes
o Password protection is active - no changes is possible
No
o Password protection is not active - changes are possible
2.02Setting Change
password
Change the password
2.03Setting Select
mode
Selection of the mode is adjusted to the measuring task.
Volume
Volumetric flow rate measurement without consideration of temperature influences.
Volume at X°
Q, QA or QB are converted with temperature and density table to mass. Then a conversion takes
place with the Volume at X° density table.
Volume at TempA
Calculates the consumption at the temperature flowmeterA. This temperature usually corresponds to
the temperature of the liquid in the reservoir.
When a mass unit is selected, the system changes automatically to the mass calculation mode.
Note:
Reference temperature X° Is freely selectable, see 2.04Setting select temperatureX
2.04Setting select tem-
peratureX
Setting the reference temperature
Requirement:
2.03Setting Select mode set to Volumes for X.
2.05Setting Select unit
rate
Set values unit rate
2.06Setting Select unit
total
Set values unit total
Operating instructions
OIE 12en-GB Edition 2020-03