11 Menu description
11.6 Menu4: K-factors flowmeterA
11.6 Menu4: K-factors flowmeterA
o Entry of the K-factors of the flowmeterA with increasing frequency for the formation of the lineariza-
tion characteristic
o K-factors, associated frequencies and resulting K-factor, see calibration certificate of the flowmeter
o Resulting K-factor, also see rating plate of the flowmeter
o Changes only possible with password
o Ä Operation,Page22
Menu item Description
4.01K-factor
VolumeterA Point1
...
4.07K-factor
VolumeterA Point7
Example without linearization:
o Point1: Resulting K-factor and frequency at Q
nom
o Point2: Frequency0
Example linearizion with three points:
o Point1: Any K-factor and associated frequency
o Point2: K-factor and associated frequency in ascending order
o Point3: K-factor and frequency at Q
nom
(highest value)
o Point4: Frequency0
Note:
The number of linearization points is limited by the input of the frequency 0.0 . Ensure that the K-
factor and frequency of Q
nom
are entered respectively in the preceding point.
The linearization is extended to 0Hz or to ∞ Hz via the first or last two linearization points resp. and
mirrored into negative values.
11.7 Menu5: K-factors flowmeterB
o Entry of the K-factors of the flowmeterB with increasing frequency for the formation of the lineariza-
tion characteristic
o K-factors, associated frequencies and resulting K-factor, see calibration certificate of the flowmeter
o Resulting K-factor, also see rating plate of the flowmeter
o Changes only possible with password
o Ä Operation,Page22
Menu item Description
5.01K-factor
VolumeterB Point1
...
5.07K-factor
VolumeterB Point7
Example without linearization:
o Point1: Resulting K-factor and frequency at Q
nom
o Point2: Frequency0
Example linearizion with three points:
o Point1: Any K-factor and associated frequency
o Point2: K-factor and associated frequency in ascending order
o Point3: K-factor and frequency at Q
nom
(highest value)
o Point4: Frequency0
Note:
The number of linearization points is limited by the input of the frequency 0.0 . Ensure that the K-
factor and frequency of Q
nom
are entered respectively in the preceding point.
The linearization is extended to 0 or to ∞Hz via the first or last two linearization points resp. and
mirrored into negative values.
Operating instructions
OIE 12en-GB Edition 2020-03
35