20
6-7
MAXMeasurement(Responsetime:400ms)
The MAX measurement mode is used to display a maximum
readingoveracertainperiodoftime.Thisfunctionisavailableon
allrangesotherthan
Ω
ranges.
#
DANGER
Nevermakemeasurementonacircuitinthefollowingcategories;
Measurementcategory
Ⅳ
(CAT.
Ⅳ
)
:overAC/DC600V
Measurementcategory
Ⅲ
(CAT.
Ⅲ
)orlower:overAC750/DCI000V
Do not make measurement with the battery compartment cover
removed.
Do not make measurement with the test leads connected to the
instrument.
Keep your fingers and hands behind the barrier during
measurement.
a.SettheFunctionSelectorswitchtothedesiredposition.
b.PresstheModebuttontoselecttheMAXmeasurementmode.
shouldbeshownonthedisplay.
c.In order to take acorrect reading, press the Zero Adjust/Reset
button once after clamping the jaws ontothe conductor or
connectingthetestleadstothecircuitundertest.
d.Thedisplayshowsthemaximumreadingduringmeasurement
e.Press the Zero Adjust/Reset button once again to return to the
normalmeasurementmode.
NOTE
◇
DataHoldfunctionisdisabledinMAXmeasurementmode.
◇
For measurement over a period more than 10 minutes, disable
the Sleep function according to the instruction in section
7-1:
SleepFunction
. Otherwise, the instrument automatically turns
itselfoffinabout10minutes.