20
6-7
MAXMeasurement(Responsetime:400ms)
The MAX measurement mode is used to display a maximum
readingoveracertainperiodoftime.Thisfunctionisavailableon
allrangesotherthan
Ω
ranges.
#
DANGER
󱾀
Nevermakemeasurementonacircuitinthefollowingcategories;
Measurementcategory
(CAT.
)

:overAC/DC600V
Measurementcategory
(CAT.
)orlower:overAC750/DCI000V
󱾀
Do not make measurement with the battery compartment cover
removed.
󱾀
Do not make measurement with the test leads connected to the
instrument.
󱾀
Keep your fingers and hands behind the barrier during
measurement.
a.SettheFunctionSelectorswitchtothedesiredposition.
b.PresstheModebuttontoselecttheMAXmeasurementmode.

shouldbeshownonthedisplay.
c.In order to take acorrect reading, press the Zero Adjust/Reset
button once after clamping the jaws ontothe conductor or
connectingthetestleadstothecircuitundertest.
d.Thedisplayshowsthemaximumreadingduringmeasurement
e.Press the Zero Adjust/Reset button once again to return to the
normalmeasurementmode.
NOTE
 ◇
DataHoldfunctionisdisabledinMAXmeasurementmode.
 ◇
For measurement over a period more than 10 minutes, disable
the Sleep function according to the instruction in section
7-1:
SleepFunction
. Otherwise, the instrument automatically turns
itselfoffinabout10minutes.