7
OUTPUTVoltage(Outputimpedance:about10k
Ω
)
Range
DCOutputVoltages
InputCurrents
Accuracy
DC
400A 0to400.0mV 0to400A
±
1.5%rdg
±
3mV
2000A 0to200.0mV 0to2000A
±
1.5%rdg
±
3mV
AC
400A 0to400.0mV 0to400A
±
1.5%rdg
±
3mV(50/60Hz)
±
3.0%rdg
±
3mV(40
〜
500Hz)
±
5.0%rdg
±
3mV(500
〜
1kHz)
2000A
0to100.0mV 0to1000A
100.1to200.0mV 1001to2000A
±
3.0%rdg
±
3mV(50/60Hz)
*Electromagneticcompatibility(EN61000-4-3)
:RFfieldstrength=<=1V/m,totalaccuracy=specifiedaccuracy
:
RFfieldstrength=3V/m,totalaccuracy=specifiedaccuracy+1%ofrange
OperatingSystem :DualIntegration
Display :Liquid crystal display with maximum
countsof4000
LowBatteryWarning :"BAIT" symbol is displayed on the digital
display.
OverrangeIndication :On each range, when the measured value
exceeds the measuring range, "OL" mark
appearsontheLCD.
ResponseTime :Approx.2seconds
Samplerate :Approx.2.5timespersecond
Locationforuse :Indoor/Outdooruse,Altitudeupto2000m
TemperatureandHumidityforGuaranteedAccuracy
:23
℃
±
5,relativehumidityupto85%
withoutcondensation
OperatingTemperatureandHumidity
:0to40
℃
,relativehumidityupto85%
withoutcondensation
StorageTemperatureandHumidity
:-20to60
℃
,relativehumidityupto90%
withoutcondensation