April 8, 2011 HN FAQ Memo, Version 2.0 Page 49 of 108
Set valves to deliver water from rinse container:
On SI parlor terminal →
On SI VMS terminal turn vacuum on and then →
|--------------------|
|4231 CIP |
|Rinse WaterIn Off<|
|Rinse WaAirOut On |
|Clean WaterIn Off |
|--------------------|
|--------------------|
|4221 MultiValve |
|Arm Position 11<|
|Lift Arm On |
|Wash Inlet Off |
|--------------------|
Start the pump
|--------------------|
|4215 Sample Outlet |
|Safety Off |
|Main Line Off |
|Milk pump Normal*|
|--------------------|
Use a flashlight to light up the dosage funnel and
see if the water is coming out of the dosage needle.
The water should come out of the needle with high
velocity, hit the slanted surface of the dosage funnel
and be deflected to the opposite side and then flow
out through the waste tube.
Stop the pump
If the water is deflected to high so it flows out of the
funnel, the setting for “home to clean” most likely
should be smaller (default is 300).
Stop the pump