April 8, 2011 HN FAQ Memo, Version 2.0 Page 50 of 108
If the water is not hitting the surface of the funnel
and just flowing directly out through the tube the
setting “home to clean” should be bigger (default is
300).
Stop the pump
HN1
If the position needs adjusting, change the setting
“home to clean” and save to repository. ±40
represent approximately ±1mm change in needle
position.
|--------------------|
|23311 Travers pos. |
|Home to Clean 300<|
|Home to Start 161 |
|Home to Lat. 920 |
|Home to Col. 2014 |
|Sensor backlash 50 |
|Save to Repository <|
|--------------------|
HN1
Initialize dosage module and check the position
again
|--------------------|
|4232. Auto cmd |
| |
| |
|42321. Init |
|--------------------|