April 8, 2011 HN FAQ Memo, Version 2.0 Page 56 of 108
8. Mount the new parts, the dosage module has new screws, but Flow module and Stop
valve had to use the old ones.
9. Connect all the wires again.
10. Connect the two new tubes from dosage module, to flow module (part nr. 60005192)
11. Connect the tube from SI again.
12. Connect the new tube from Flow module to stop valve (part nr. 60026876)
13. Connect the new tube from Diluents bag to flow module (part nr. 60005193)
14. Connect the tube from stop valve to drain again.
15. Make sure all tubes and wires is fit correct, and then put 6 new strips on again, just to
hold, but not to tight so it damage the tubes.
16. Take new diluents, and connect it.
17. Make a Diluents prime whit the AI menu 1-Maintaince -> 2-Relpace Diluents. And
watch if there are any leaks… after a complete replace, 3-4 primes is needed to make
sure all air is out of the system again.
18. Make a cleaning whit the AI menu 1-Maintaince -> 4-Request clean , and watch when
the water comes trough, that there is no leaks.
19. If there is no leaks and all is well fitted, close the analyzer, restart the whole system whit
taking the power, and make sure the Analyser goes to standby mode.
20. Load Drybags, and when the AI is ready, load sticks and change the Drybags an extra
time.
21. Then HN is ready again.
What is in the package from Lattec :
And some new Diluent.
Check AI for leaks on Diluent supply.
Problem :
LDH, Urea, and BHB is working whit no problems, but a Progsteron is missing control
line.
Reason to the error :
No diluents / not enough mixed whit the sample on the progesterone stick.
Possible errors: