April 8, 2011 HN FAQ Memo, Version 2.0 Page 57 of 108
1. There is a leak on the diluents supply line between the flow module and dosage
module.
2. No O-rings on the Needle, or the needle is not put on correctly.
Solution on the problem :
Problem 1 :
1. Open AI, as described on the service Memo.
2. When the AI is open, start a Diluents prime : AI menu :
a. 4 Service -> 2 Simple func -> 3 Dosage -> 2 Auto cmd -> 2 prime.
b. If it is needed, choice 1. Init first instead of 2. prime.
3. When priming, watch if there is any leaks on the tube between Flow module, and
Dosage module (Se pictures).
4. If there is a leak, change the tube on pictures :
Problem 2 :
1. Take out needle.
2. Check if both O-rings is there.
3. Put in Needle incl. O-rings again.