EasyManuals Logo

LCR Elite2 User Manual

Default Icon
26 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
19 User Manual - LCR Elite2 - ver 1.3
Resistance Accuracy Specifications
Table 4-5 LCR Elite2 Resistance Accuracy Specification
Range
Resolution
Accuracy = A
Z
+ Offset
100Hz
120Hz
1kHz
10kHz
1000mΩ
0.1m
0.5% + 20m
0.5% + 20m
0.5% + 20m
0.5% + 20m
10Ω
0.001
0.5% + 0.02
0.5% + 0.02
0.5% + 0.02
0.5% + 0.02
100Ω
0.01
0.3% + 0.03
0.3% + 0.03
0.3% + 0.03
0.3% + 0.03
1000Ω
0.1
0.3% + 0.3
0.3% + 0.3
0.3% + 0.3
0.3% + 0.3
10kΩ
0.001k
0.3% + 0.003k
0.3% + 0.003k
0.2% + 0.003k
0.3% + 0.003k
100kΩ
0.01k
0.3% + 0.05k
0.3% + 0.05k
0.2% + 0.05k
0.3% + 0.05k
1000kΩ
0.1k
1.0% + 0.5k
1.0% + 0.5k
0.3% + 0.5k
1.0% + 0.5k
10MΩ
0.001M
2.5% + 0.008M
2.5% + 0.008M
2.5% + 0.008M
---
Inductance Accuracy Specifications
Table 4-6 LCR Elite2 Inductance Accuracy Specifications
Range
Resolution
Accuracy = A
L
+ Offset
100Hz
120Hz
1kHz
10kHz
1000nH
0.1nH
---
---
---
1.5% + 0.1uH
10uH
0.001uH
---
---
---
1.5% + 0.1uH
100uH
0.01uH
---
---
---
0.7% + 0.1uH
1000uH
0.1uH
---
---
0.5% + 0.3uH
0.7% + 0.3uH
10mH
0.001mH
---
---
0.4% + 0.003mH
0.5% + 0.003mH
100mH
0.01mH
0.5% + 0.03mH
0.5% + 0.03mH
0.4% + 0.03mH
0.7% + 0.03mH
1000mH
0.1mH
0.4% + 0.3mH
0.4% + 0.3mH
0.7% + 0.3mH
---

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LCR Elite2 and is the answer not in the manual?

LCR Elite2 Specifications

General IconGeneral
BrandLCR
ModelElite2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals