19 User Manual - LCR Elite2 - ver 1.3
Resistance Accuracy Specifications
Table 4-5 LCR Elite2 Resistance Accuracy Specification
Range
Resolution
Accuracy = A
Z
+ Offset
100Hz
120Hz
1kHz
10kHz
1000mΩ
0.1m
0.5% + 20m
0.5% + 20m
0.5% + 20m
0.5% + 20m
10Ω
0.001
0.5% + 0.02
0.5% + 0.02
0.5% + 0.02
0.5% + 0.02
100Ω
0.01
0.3% + 0.03
0.3% + 0.03
0.3% + 0.03
0.3% + 0.03
1000Ω
0.1
0.3% + 0.3
0.3% + 0.3
0.3% + 0.3
0.3% + 0.3
10kΩ
0.001k
0.3% + 0.003k
0.3% + 0.003k
0.2% + 0.003k
0.3% + 0.003k
100kΩ
0.01k
0.3% + 0.05k
0.3% + 0.05k
0.2% + 0.05k
0.3% + 0.05k
1000kΩ
0.1k
1.0% + 0.5k
1.0% + 0.5k
0.3% + 0.5k
1.0% + 0.5k
10MΩ
0.001M
2.5% + 0.008M
2.5% + 0.008M
2.5% + 0.008M
---
Inductance Accuracy Specifications
Table 4-6 LCR Elite2 Inductance Accuracy Specifications
Range
Resolution
Accuracy = A
L
+ Offset
100Hz
120Hz
1kHz
10kHz
1000nH
0.1nH
---
---
---
1.5% + 0.1uH
10uH
0.001uH
---
---
---
1.5% + 0.1uH
100uH
0.01uH
---
---
---
0.7% + 0.1uH
1000uH
0.1uH
---
---
0.5% + 0.3uH
0.7% + 0.3uH
10mH
0.001mH
---
---
0.4% + 0.003mH
0.5% + 0.003mH
100mH
0.01mH
0.5% + 0.03mH
0.5% + 0.03mH
0.4% + 0.03mH
0.7% + 0.03mH
1000mH
0.1mH
0.4% + 0.3mH
0.4% + 0.3mH
0.7% + 0.3mH
---