3–1
3
Installation and Safety
Installation for Safe and Efficient
Operation
The oscilloscope will operate to its specifications if the
operating environment is maintained within the following
parameters:
Ø Temperature.......................... 5 to 40 °C (41 to 104 °F) rated.
Ø Humidity................................Maximum relative humidity 80 % RH
(non-condensing) for temperatures up
to 31 °C decreasing linearly to 50 %
relative humidity at 40 °C
Ø Altitude ..................................< 2000 m (6560 ft)
The oscilloscope has been qualified to the following EN61010-1
category:
Ø Protection Class .........................................I
Ø Installation (Overvoltage) Category ...........II
Ø Pollution Degree.........................................2
Safety Symbols Where the following symbols or indications appear on the
instrument’s front or rear panels, or elsewhere in this manual, they
alert the user to an aspect of safety.
Symbol Meaning
CAUTION: Refer to accompanying documents (for Safety-
related information).
See elsewhere in this manual wherever the symbol is present,
as indicated in the Table of Contents.
CAUTION: Risk of electric shock.
x
On (Supply).
Operating Environment