System Commands 5
STATUS *OPC
Command/Query
DESCRIPTION
COMMAND SYNTAX
The *OPC (OPeration Complete) command sets the OPC bit (bit
0) in the standard Event Status Register (ESR) to true. This com-
mand has no other effect on the operation of the oscilloscope as
the instrument starts parsing a command or query only after it has
completely processed the previous command or query.
The *OPC? query always responds with the ASCII character "1"
as the oscilloscope responds to the query only once the previous
command has been entirely executed.
Note: This command can be executed in both local and remote
modes.
* OPC
QUERY SYNTAX
Response format
RELATED COMMANDS
*OPC?
*OPC 1
* WAI