System Commands
|
1
Value
21
22
23
24
25
26
30
31
32
33
34
35
50
51
52
53
54
55
56
57
58
59
60
61
62
64
Description
Permission error. The command cannot be executed In local mode.
Environment error. The instrument is not configured to correctly process a command.
For instance, the oscilloscope cannot be set to RIS at a slow time base.
Option error. The command applies to an option which has not been installed.
Unresolved parsing error.
Parameter error. Too many parameters specified.
Non-implemented command.
Hex data error. A non-hexadecimal character has been detected in a hex data block.
Waveform error. The amount of data received does not correspond to descriptor indi-
cators.
Waveform descriptor error. An Invalid waveform descriptor has been detected.
Waveform time error. Invalid RIS or TRIG time data has been detected.
Waveform data error, invalid waveform data have been detected.
Panel setup error. An Invalid panel setup data block has been detected.
No memory card present when user attempted to access the card.~
Memory card not formatted when user attempted to access the card.*
Memory card was exchanged when user attempted to RECALL the NEXT, PREVIOUS
or SAME waveform.~,
Memory card was write protected when user attempted to create a file. to delete a
file. or to format the ¢ard.~,
Bad memory card detected during formatting.
Memory card root directory full. LECROY_I .DIR subdlrectory cannot be created.*
Memory card full when user attempted to write to It.*
Memory card file sequence numbers exhausted (999 reached).*
Memory card file not found.*
Attempt to retrieve a file from memory card that Is neither a waveform nor a panel.*
Memory card file Is write protected (DOS Read/Only attribute).*
Memory card fllename not DOS compatible, or waveform fllename begins with an
"A" .,r
A retrieved panel or waveform contains invalid bytes, or a waveform Is too long for the
Instrument.~,
Read/Write outside of memory card cluster chains. Can Indicate a corrupted FAT
table.
Can also Indicate a waveform or panel file with a length shorter than the Internal byte
count.
84
Execution Error Status Register Structure (EXR)
Table 6
only oscilloscopes equipped with the memory card option