NOVOLAS Basic AT Version 2.1
09/2010
42 Operation
Register
Index
Name Function Units
Value range
6 TSet Target temperature
- This sets the target temperature for
corresponding closed loop control modes.
The unit depends on bit 0 of register 1.
7 ISet Sets the current if
digital current set
is used.
0.1A 0-2000, preset 0
(ex: 100= 10A, 200=20A)
(If the air cooled laser module is used, the
unit is 0.01 A)
8,
10,12,..
P_IX Sets the current of
an interval in
profile mode
0.1A 0-2000, 16 Register P_IO-P_l15
(if the air cooled laser is used, the unit is
0.01A)
9, 11,
13
P_TX Sets the duration
of an interval in
profil mode.
1 ms
0-60000, 16 Register P_TO-P_T15
Register
Index
Name Function Units
Value range
900 Guiding Laser If a 1 is written, the
pilot laser will be
switched on.
Note: if the Input
“Guidung laser off” is
activated on the
Customer IO, the
guiding laser will stay
inactive.
- 1 (default): guiding laser will be on.
else: guiding laser will be switched off
999 Store
Parameters
If a 1 is written the
parameters of the
selected laser
module are stored
permanently
-
0:
The following commands are applied to
laser unit #A
1:
The following commands are applied to
laser unit #B
2:
The following commands are applied to
laser unit #C
9900 Select
Laser Unit
Defines the laser
unit, on which the
following
parameters are
applied
-
3:
The following commands are applied to
laser unit #D