Page #12 background imageLoading...
Page #12 background image
User Manual for PuraLev
®
600MU
www.levitronix.com
PL-4033-00, Rev09, DCO# 20-144
12
Cable
Jacket
Cable OD
Motor Sensor
Cable OD
Motor Power
Minimum Bending Radius
Permanent Installation
Minimum Bending Radius
Sometimes Moved
FEP
6.60 mm
8.40 mm
7x Cable OD
15 x Cable OD
PVC
7.20 mm
11.0 mm
5x Cable OD
11 x Cable OD
Table 6: Specifications for minimum bending radius of motor and adaptor cables
Note: If not mentioned explicitly all the cables are not suitable for constant dynamic bending and movement!
Figure 11: Basic dimensions (in mm and [inch]) of air cooling module ACM-600.2
Note 1: Non-tolerated dimensions are for reference only. Note 2: Not to be used for ATEX / IECEx applications)
Figure 12: Basic dimensions (in mm) of controller LPC-600.2
Note 1: Non-tolerated dimensions are for reference only. Note 2: Same basic dimensions for LPC-600.1)
Alternative for
mounting DIN
rail clips
Screw Torque Spec.:
M3: 20 Ncm (Stainless Steel)
M5: 40 Ncm (Stainless Steel)