Page 153 of 192
The HB50 is primarily designed for the LINAK JUMBO system and LINAK
control boxes with memory functions.
The HB50 gives the user access to a range of memory functions, allowing
present positions to be stored.
For use in a wide variety of hospital and patient care.
17. HB50 (MEDLINE
®
CARELINE
®
)
The HB60 series is exclusively designed to be used together with the LINAK
HOMELINE
®
system: LA31/LA29 and CB9 with or without memory.
18. HB60 (HOMELINE
®
)
Usage:
• Ambient temperature: +5 °C to +40 °C
• HB50B is compatible with CBJ2
• HB50E is compatible with CB09P..Px types (except ..PM / PN) and CB14 with memory
• HB50J is compatible with CBJ1
• HB50Y is for Jumbo Home (CBJH)
LINAK offers the HB70 with protection class IPX6 as standard and a
range of options such as control of up to 5 actuators, memory, and
simultaneous drive. The HB70 is designed to operate with most LINAK
control boxes.
Usage:
• Compatible with most LINAK control boxes.
• Approved according to: EN 60601-1, EN 60335-1 and UL 60601-1 as part of a LINAK actuator system
Recommendation
• It is not possible to combine HB7x with the binary based CB9..PM/PN.
• The IPX6 Washable version has a special adhesive for the front covers.
• The HB75xE0 used together with CB140 will give trend and anti-trend on channel 1 and 2 of the control box when using the last button row.
• All front covers use the codes W0 (not Washable) and WW (Washable) Memory:
Memory:
• The memory and parallel functions require the control box to have a microprocessor.
• When storing a memory position on the control box, the actuators must run to the desired position and the “store” button (S) must be pushed.
Then the desired memory position button (1, 2 or 3) must be activated within 2 seconds.
19. HB70 (MEDLINE
®
CARELINE
®
)