17. QLCI2 (MEDLINE
®
CARELINE
®
)
The Quad Load Cell Interface 2 (QLCI2) for the hospital and care segment is a scale system accessory with weighing capabilities and Out of Bed
functionality.
The housing, makes the QLCI2 easy to mount by unique slide-on brackets and has an IPX6 Washable DURA
TM
ingress protection.
The QLCI2 supports the LINAK OpenBus
TM
system offering a high level of customisation.
LINAK A/S delivers the OpenBus™ system only and is not responsible for any products (i.e. products from 3
rd
party suppliers) other than
LINAK products or the compatibility of such products with the LINAK OpenBus™ system.
Usage
Operation temperature: + 5 °C to + 40 °C
Storage temperature: - 10 °C to + 50 °C
Relative humidity: 20% to 80% - non-condensing
Atmospheric pressure: 700 to 1060 hPa
Approvals IEC60601-1
ANSI/AAMI ES60601-1
CAN/CSA-22.2 No 60601-1
EN 45501 / OIML R76
EU type examination according to 2014/31/U
Compatibility: All OpenBus
control boxes
Flammability rating: UL V0
Latex free: Yes
Approvals
An OEM application approval according to EN45501
Typically the LINAK Weighing Solution will be classified as a “non-automatic weight system.
If a LINAK LINAK Weighing Solution system is applied to a bed AND classified as such, the system MUST be first-time verified and sealed.
The verification and sealing is typically carried out in one of two ways:
1. Verification by the Bed manufacturer himself.
It prescribes that the manufacturer is certified to carry out the verification.
The certification can be obtained through a Notified Body that performs auditing and approval of the procedures and the quality system in the
manufacturing company.
An example from Denmark:
’DS Certificering’ is the only Notified Body in Denmark, certified to carry out approvals of quality systems for manufacturing and calibration of
‘non-automatic weight’ systems. Within Europe it is however possible to use any other Notified Body from one of the EU member states.
When certified the Bed manufacturer obtains a type approval certificate to prove they are certified to manufacture and calibrate their own “non-
automatic weight system.
2. Verification by ”first-time” verification Bodies.
In Denmark there are three Notified Bodies available for the verification and sealing of the application:
Force Technology, Dansk Kalibreringsteknik and Trescal.
Again any other Notified Body from an EU member state can be used.
First-time verification can take place at either the manufacturer or at the destination of use.
Requirements in both situations:
The Type Approval Certificate number MUST be marked at the label on the weight unit.
The Type Approval Certificate must be issued according to and including reference to the Directive for “non-automatic weights 2009/23/EC
(new non-modified version of 90/384/EEC).