5.3 Manual setting of the CPR
The CPR controller is marked with red information labels on both sides of the back section. The labels are
marked with CPR text.
To activate the function, use one hand to push the red lever in the specified direction. Use your other
hand to pull down the back section by the handle on the back section. Keep holding the lever during
this movement.
To reset the back section to its original position, use motoric positioning (activates via hand
control, side control panel or side rail panel).