8.5 Storage instruction
Packed birthing beds may be exposed to conditions not exceeding the following limits for the period of
up to 15 weeks:
ambient temperature (-10° to + 50°)
relative air humidity (from 10 % to 95 %)
regional air pressure (500 hPa to 1060 hPa)
8.6 Ambient conditions for operation
ambient temperature (10° to + 40°)
relative air humidity (from 30 % to 75 %)
regional air pressure (700 hPa to 1060 hPa)
8.7 Service and repairs
Warning:
Caution
All service revisions, safety and technical inspections or service interventions must be executed only by a
technician trained by LINET company.
Diagrams, lists of product parts, descriptions or other information intended to help the service personnel
with the repair of product parts, repairable by the service personnel only as implied by the manufacturer,
are available upon demand at the manufacturer.
The product is provided with the warranty period of 2 years on conditions that regular annual service
inspections are carried out.
For all medical equipment produced by the LINET s.r.o company, the operator must ensure regular
technical safety inspections according to the producer’s recommendation, thus the inspections must take
place at least 1x per year or after the carrying out of any repair of the medical equipment or after any
intervention regarding the electrical system of the medical equipment. The details about the scope,
parameters and procedures of the technical safety inspection are specified in the service manual of the
medical equipment.
The extended warranty does not cover:
batteries (these are provided with the warranty of 6 months)
From practical testing results, the technical lifespan of the product is 15 years; lifespan of the battery is
limited and must be replaced; gas springs need to be replaced every 12 years.
In case the product is not functioning, please contact the product dealer in your country. In case of any
other problems, please contact (preferably in English) the manufacturer at:
For repairs, please use only the original LINET spare parts and the services of trained service technicians
licensed to repair products of this type. Unprofessional repairs shall be considered as a substantial breach
of the warranty terms and conditions and the manufacturer simultaneously does not bear any
responsibility for any further possible damage to the product which might result from unprofessional
intervention.