10.3 Sprung Retractable Fifth Castor (optional)
During middle position of pedals Sprung Retractable Fifth Castor is around 12 mm above oor.
When bed is braked Sprung Retractable Fifth castor is around 65 mm above oor.
To activate the Fifth castor:
Press green Drive pedal to the lower position.
To retract the Fifth castor:
Leave all Brake pedals and Drive pedals unpressed in their middle position.
10.4 Angle Indicators (optional)
Angle Indicators are optionally situated on both sides of the Backrest or on both sides of the Backrest and on both sides of Mattress
Support Platform frame in seat section. Backrest angle indicators are intended for an approximate Backrest angle measurement.
Angle indicators in the seat section are intended for an approximate measurement of Trendelenburg tilt and Anti-Trendelenburg tilt.
Fig. Angle Indicators