13 Cleaning/Disinfection
WARNING!
Risk of injury due to accidental bed movement!
Ensure pedals will not be pressed accidentally during cleaning.
WARNING!
If oil leaks from hydraulic units or gas springs contact the manufacturer´s service department!
CAUTION!
Material damage due to incorrect cleaning/disinfection!
Do not use washing tunnels to clean the bed.
Do not use pressure or steam cleaners.
Follow the instructions and observe the dosages recommended by the manufacturer.
Ensure that disinfectants are selected and applied exclusively by qualied hygiene experts.
Respect used materials during cleaning and desinfection! For information see the following table.
Bed components Materials used
Head, Foot and Side Covers Polypropylene
Siderails Powder Coated Steel + Polypropylene
Mattress Support Platform Covers HPL
Castors Polyurethane + Steel + Polyamide
Frame structure Powder Coated Steel
Column Cover PVC
Undercarriage Cover ABS
Corner Covers Polypropylene
Corner Bumpers Polypropylene
Siderail Release Handles Powder Coated Steel + PVC
For safe and gentle cleaning:
Do not use any strong acids or bases (optimum pH range 6 - 8).
Exclusively use detergents that are suitable for cleaning medical equipment.
Do not use abrasive powders, steel wool, or other materials and cleaning agents that might damage the mattress replace
ment system.
Never use any corrosive or caustic detergents.
Never use detergents that deposit calcium carbonate.
Never use detergents with solvents that might affect the structure and consistency of the plastics (benzene, toluene,
acetone, etc.).
Observe local directives regarding infection control.
Make sure any cleaning agent used is approved by:
   ■ the facility in which the mattress replacement system is to be used.
   ■ by the EPA (Environmental Protection Agency) of the country in which the mattress replacement system is to be used.
13.1 Cleaning (Sprint 100)
Prepare for cleaning as follows:
Put the Mattress Support Platform in the highest position.
Adjust the Backrest and Thighrest so that the reverse sides are accessible.
Move the bed to the location where it will be cleaned.
Lock the brakes on the bed.

Table of Contents

Other manuals for LINET Sprint 100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET Sprint 100 and is the answer not in the manual?

LINET Sprint 100 Specifications

General IconGeneral
BrandLINET
ModelSprint 100
CategoryMedical Equipment
LanguageEnglish

Related product manuals