TR700 Instruction Manual
41
V
0
Output signal of load cell when weighing system is at no load;.
V
1
Output signal of load cell when weighing system is at a certain load;
I
0
The display value of V
0
input to the meter without calibration;
I
1
The display value of V
0
input to the meter without calibration;
W
0
The display value of V
0
input to the meter after calibration. (I.e. zero);
W
1
The display value of V
0
input to the meter after calibration. (I.e. the standard force values which
correspond to the scale division or load).
Note during the calibration:
W
1
is
no less than 100 divisions;
Resolution
dV
WW
VV
/5.0
01
01
µ
;
The range of output signal of Strain gauge load cell : approx. 030mV;
Calibration is to realize the new corresponding relation;
The input signal of load cell of transmitter should meet0.05mV≤V
0
≤15mV,V
0
V
1
≤32mV.
7.2. Operation of the Calibration
7.2.1. Steps
In the normal statepress to enter the menuand press to choose“CAL” and press
to display“CA L1”; and press to choose“CAL1“CAL2”or“CAL3”and press
to enter calibration parameter setting.