V
0
:Output signal of load cell when weighing system is at no load;.
V
1
:Output signal of load cell when weighing system is at a certain load;
I
0
:The display value of V
0
input to the meter without calibration;
I
1
:The display value of V
0
input to the meter without calibration;
W
0
:The display value of V
0
input to the meter after calibration. (I.e. zero);
W
1
:The display value of V
0
input to the meter after calibration. (I.e. the standard force values which
correspond to the scale division or load).
Note during the calibration:
① W
1
is
no less than 100 divisions;
② Resolution :
;
③ The range of output signal of Strain gauge load cell : approx. 0~30mV;
Calibration is to realize the new corresponding relation;
④ The input signal of load cell of transmitter should meet:0.05mV≤V
0
≤15mV,V
0
<V
1
≤32mV.
7.2. Operation of the Calibration
7.2.1. Steps
In the normal state,press to enter the menu,and press 、 to choose“CAL” ,and press
to display“CA L1”; and press 、 to choose“CAL1”、“CAL2”or“CAL3”,and press
to enter calibration parameter setting.