TR700 Instruction Manual
60
11. Attach
11.1. Function Table
11.1.1. Parameter List of General Function Setting
Number
F XX
Name
Defaul
t
Setting
Paramet
ers
Range
Description
00
Zero range
3
0-10
0 Zero function off;
1-10 It is 1-10 of the capacity.
01
Zero tracking range
0
0-10
The display must be in zero tracking range
in 2s, and later it goes back to zero.
0 Do not perform zero tracking;
1-10
Display division.
02
Motion detection range
3
0-10
0 Motion detection is off;
1-10 Display division.
03
Filter coefficient
3
0-9
0 No filtering;
1-9 The larger the figure is, the greater
the filter is.
04
Display updating rate
3
0-4
0
1 times per second
1 4 times per second
2
8 times per second
3
16 times per second
4
30 times per second
05
Unit conversion
0
0-1
0
kg
1
t
06
Sampling rate
0
0-3
0
25 times per second
1
50 times per second
2
100 times per second
3
200 times per second
07
Conditions of
comparison output
0
0-2
0
Gross weight
1
Net weight
2
Display value
08
Distribution of external
control input 1
3
0-8
0
Gross/net
1
Tare
2
Clear zero
3-7
Extended Functions
8
Keyboard lock
only lock the
“zero,
Tare and gr/nt” keys
0
Gross/net
1
Tare