capacity; press to save the capacity input
(
If the sensitivity input setting is jumped
over
,
the capacity will not be saved.
)
and get back to the “CAL2” screen
,
or press to
get back to the “CAL2” screen directly without saving the capacity input.
7.2.4. Modification of the Calibration Coefficient
“CAL3”is displayed, and press to enter the first step.
Step 1:Input the Password
“000000”is displayed,and press to enter the password input setting; press 、 、 to
input the password and enter the next step.
Step 2: Input the Calibration Coefficient
Press
、 、
to input the calibration coefficient; press to get back to “CAL3”screen
without saving the calibration coefficient input; or press to enter the third step.
Step 3: Save the Calibration Coefficient
“SAVEr”is displayed, cluing to save the calibration coefficient input or not; press to get back
to “CAL3”screen without saving the calibration coefficient, or press to save it.
7.3. CAL Drawing
Table 7-1