TR700 Instruction Manual
43
division input and enter the third step, or press to enter the third step directly without
saving the division input.
Step 3Capacity Setting
C1.MAX”is displayed, and press to jump over the capacity setting; press to
enter the capacity setting; press
to input the capacity; press to save
the capacity input and enter the forth step, or press to enter the forth step directly
without saving the capacity input.
Step 4Zero Calibration
“C1.Zero” is displayed, and press to jump over the zero setting; press to enter
the zero calibration, and “000000”is displayed and all are flickering. If the dynamic
detection is on, after the scale is stable and the dynamic indicator is off, press to
perform zero calibration and enter the fifth step, or enter the fifth step without zero
calibration.
Step4Weighing Calibration
“C1.SPAn” is displayed
and press to jump over the weighing calibration; press
to enter the weighing calibration; press
to input the weight. If the
dynamic detection is on, after the scale is stable and the dynamic indicator is off, press
to save the weight input and get back to the “CALL” screen, or press get back
to the “CALL” screen directly without saving the weight input.