EasyManuals Logo

Maico easyScreen User Manual

Maico easyScreen
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
Operation Manual easyScreen
8508935 Rev. 4
35
04/04/2019
Table 4 BERAphone® LED Behavior
Phase
Status
LED Appearance
Impedance test
Good impedance
Poor impedance vertex
Poor impedance all electrodes
Testing
Test running quality OK
Artifacts occurring
Paused
Electrode off
Test complete
Establishing a PC-Connection
To transfer data to a PC, establishing a PC-connection via USB is required. If the
easyScreen is used with office equipment that is not a medical device itself (see Table
5, PC-Connection 1), make sure to establish the PC-connection in one of the following
ways (see Table 5, PC Connection 2, 3 or 4).
Make sure you use only office equipment with the device
that is a medical device itself or meets the requirements of
IEC 60950. If a non-medical device is used within the patient
environment (1.5 m from patient as defined in IEC 60601) a
voltage transformer must be used (exception: a battery
driven laptop is used).
WARNING

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Maico easyScreen and is the answer not in the manual?

Maico easyScreen Specifications

General IconGeneral
BrandMaico
ModeleasyScreen
CategoryMedical Equipment
LanguageEnglish

Related product manuals