EasyManua.ls Logo

Masimo SET Rainbow SET Rad-57 - Initial Monitor Setup

Masimo SET Rainbow SET Rad-57
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-2
Rad-57 Signal Extraction Pulse CO-Oximeter Operator’s Manual
3
Setup
Monitor Setup
Initial Setup
1. Inspect the oximeter case for damage.
2. Install 4 (four) new AA alkaline batteries.
3. Turn the Instrument on, the LEDs will scroll in the display window as the sensor calibrates,
verify all indicators illuminate and speaker sounds a brief tone.
4. Configure the Instrument for your regional power line frequency (50 or 60 Hz) if needed.
Default is 60 Hz (standard for the United States). See Section 4, Special Menu, Special menu
- Line Frequency Configuration.
CAUTION: THE INSTRUMENT MUST BE CONFIGURED TO MATCH YOUR LOCAL POWER LINE
FREQUENCY TO ALLOW FOR THE CANCELLATION OF NOISE INTRODUCED BY FLUORES-
CENT LIGHTS AND OTHER SOURCES.
No other setup is required. Refer to Section 4, General Setup and Use for additional steps to verify
proper functioning of the Instrument.

Table of Contents

Related product manuals