Rad-57 Signal Extraction Pulse CO-Oximeter Operator’s Manual
8
M-LNCS
/LNCS
®
Specialty Sensors
M-LNCS sensors must be used in conjuction with Rainbow patient cable series.
LNCS sensors must be used in conjunction with Red LNC cables.
SENSOR
Weight
Range
Saturation Accuracy Pulse Rate Accuracy Low Perfusion Accuracy
No Motion Motion No Motion Motion Saturation Pulse Rate
M-LNCS/LNCS
Newborn
Infant/Pediatric
< 3 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
M-LNCS/LNCS
Newborn
Neonatal
< 3 kg ± 3% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm
M-LNCS/LNCS
Trauma
> 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOPv
TM
Adhesive Sensors
(LNOPv sensors must be used in conjunction with Red PC cables)
Weight Saturation Accuracy Pulse Rate Accuracy Low Perfusion Accuracy
SENSOR Range No Motion Motion No Motion Motion Saturation Pulse Rate
LNOPv Ad > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
Sensor Accuracy
Refer to Section 7, Specifications for SpO
2
, SpCO, SpMet and pulse rate accuracy, unless otherwise
specified in the tables above. Accuracy specified when used with Masimo Rainbow SET technology
pulse CO-Oximetry monitors or with licensed Masimo SET pulse oximetry modules during no motion.
Numbers represent ± 1 standard deviation. Plus or minus one standard deviation represents 68%
of the population. SpO
2
accuracy from 70% to 100%. Pulse Rate accuracy from 25 to 240bpm.
Carboxyhemoglobin accuracy (SpCO) from 1 to 40%. Methemoglobin accuracy (SpMet) from 1 to
15%.
Cleaning And Reuse Of Masimo Reusable Sensors and Cables
Reusable sensors and patient cables can be cleaned per the following procedure:
Remove the sensor from the patient.
Disconnect the sensor from the patient cable.
Disconnect the patient cable from the monitor.
Wipe the entire sensor and/or patient cable clean with a 70% isopropyl alcohol pad.
Allow to air dry thoroughly before returning it to operation.
CAUTION: CAREFULLY ROUTE PATIENT CABLES TO REDUCE THE POSSIBILITY OF PATIENT
ENTANGLEMENT OR STRANGULATION.
Reattachment of Single Use Adhesive Sensors
Single use sensors may be reapplied to the same patient if the emitter and detector windows
are clear and the adhesive still adheres to the skin.
NOTE: If the sensor fails to track the pulse consistently, the sensors may be incorrectly
positioned. Reposition the sensor or choose a different monitoring site.
CAUTION: DO NOT ATTEMPT TO REPROCESS, RECONDITION OR RECYCLE ANY SENSORS
AS THESE PROCESSES MAY DAMAGE THE ELECTRICAL COMPONENTS, POTENTIALLY
LEADING TO PATIENT HARM.