PERFORMANCE
Measurement Range
SpO
2
: 0 -100%
SpMet*: 0 - 99.9%
SpCO*: 0 - 99%
Pulse Rate: 25 - 240 (bpm)
Perfusion Index: 0.02% - 20%
Pleth Variability Index (PVI*): 0 - 100%
Accuracy:
Arterial Oxygen Saturation Accuracy
1
Saturation 60% to 80%
No Motion
Adults, Infants, Pediatrics ±3%
Saturation 70% to 100%
No Motion
2
Adults, Infants, Pediatrics ± 2%
Neonates
± 3%
Motion
3
Adults, Infants, Pediatrics, Neonates
± 3%
Low Perfusion
4
Adults, Infants, Pediatrics, Neonates ± 2%
Pulse Rate Accuracy
5
Pulse Rate: 25 - 240 (bpm)
No Motion
Adults, Infants, Pediatrics, Neonates
± 3 bpm
Motion
4
Adults, Infants, Pediatrics, Neonates
± 5 bpm
Low Perfusion
Adults, Infants, Pediatrics, Neonates
± 3 bpm
Carboxyhemoglobin saturation accuracy (%SpCO*)
1
Adults, Infants, Pediatrics
1% - 40% ± 3%
Methemoglobin saturation accuracy (%SpMet*)
1
Adults, Infants, Pediatrics, Neonates 1% - 15% ± 1%
Resolution
Arterial Oxygen Saturation (%SpO
2
) 1%
Carboxyhemoglobin Saturation (%SpCO*) 1%
Methemoglobin Saturation (%SpMet*) 0 .1%
Pulse Rate (bpm) 1 bpm

Table of Contents