Rad-57 Signal Extraction Pulse CO-Oximeter Operator’s Manual
4-3
4
Operation
Default Settings
The Rad-57 is shipped configured with all values set to factory default. Adjustments made by the user
will be retained after a power cycle for all values except alarm silence.
NOTE: Before use, confirm that the instrument settings are set appropriately.
The following table outlines the default values:
OPTION
FACTORY DEFAULT
SETTING
CONFIGURABLE SETTING/BEHAVIOR AT
POWER-UP
SpCO* low alarm limit (Default is Off)
Set to Pre-Power Down Setting
See Section 5, Alarm Limits for all settings.
SpCO* high alarm limit (Default is 10%)
SpMet* low alarm limit (Default is Off)
Set to Pre-Power Down Setting
See Section 5, Alarm Limits for all settings.
SpMet* high alarm limit (Default is 3%)
SpO
2
low alarm limit (Default is 90%)
Set to Pre-Power Down Setting
See Section 5, Alarm Limits for all settings.
SpO
2
high alarm limit (Default is OFF)
Pulse rate low alarm limit (Default is 50 BPM)
Set to Pre-Power Down Setting
See Section 5, Alarm Limits for all settings.
Pulse rate high alarm limit (Default is 140 BPM)
PVI
* low alarm limit
(Default is "---" Off)
Set to Pre-Power Down Setting
See Section 5, Alarm Limits for all settings.
PVI* high alarm limit
(Default is "---" Off)
PI low alarm limit (Default is "---" Off)
Set to Pre-Power Down Setting
See Section 5, Alarm Limits for all settings.
0.04 to 19 then "---"
PI high alarm limit (Default is "---" Off)
Sensitivity (Default is Normal)
Set to Pre-Power Down Setting
Normal, Maximum (HI) or APOD
Averaging Time (Default is 8 seconds)
Set to Pre-Power Down Setting
2, 4, 8, 10, 12, 14, or 16 seconds
FastSat (Default is Off)
Set to Pre-Power Down Setting
Off/On
SmartTone (Default is Off)
Set to Pre-Power Down Setting
Off/On
Display Low SIQ Readings (Default is No)
Set to Pre-Power Down Setting
Yes/No
PVI Averaging (Default is Long)
Set to Pre-Power Down Setting
Long/Short
Display brightness (Default is level 2)
Set to Pre-Power Down Setting
Levels 1 through 4
Pulse tone volume (Default is level 1)
Set to Pre-Power Down Setting
Levels Off through 3