4-18
Rad-57 Signal Extraction Pulse CO-Oximeter Operator’s Manual
4
Exception
Messages
The exceptions are displayed as a 3 digit, ASCII encoded, hexadecimal value. The binary bits of
the hexadecimal value are encoded as follows:
000 = Normal operation; no exceptions
001 = No Sensor
002 = Defective Sensor
004 = Low Perfusion
008 = Pulse Search
010 = Interference
020 = Sensor Off
040 = Ambient Light
080 = Unrecognized Sensor
100 = reserved
200 = reserved
400 = Low Signal IQ
800 = Masimo SET. This flag means the algorithm is running in full SET mode. It requires a
Masimo sensor and needs to acquire some clean data for this flag to be set.