NomoLine ISA CO2 Chapter 4: Troubleshooting
www.masimo.com 20 Masimo
Symptom Possible Cause Correction
Gas readings are
questionable
Incorrect sampling line
used
Sampling line attached
incorrectly
Sampling line leaking
Internal error
Use the correct sampling line with
NomoLine ISA CO2.
Disconnect and reconnect the
sampling line to NomoLine ISA
CO2.
Follow the Directions for Use to
properly connect the sampling
line.
Check sampling line for leaks.
Replace sampling line. See
Sampling Line Replacement on
page 14.
If persistent, the NomoLine ISA
CO2 requires service. See Service
and Return Procedure.