EasyManuals Logo

Masimo NomoLine ISA CO2 User Manual

Masimo NomoLine ISA CO2
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
www.masimo.com 19 Masimo
Chapter 4: Troubleshooting
Troubleshooting NomoLine ISA CO2
Symptom Possible Cause Correction
NomoLine ISA
CO2 values are
not displayed on
host monitor
NomoLine ISA CO2
requires a power cycle
Host device is not
configured correctly or is
incompatible with
NomoLine ISA CO2
Disconnect and reconnect the
NomoLine ISA CO2 connector at
the host device.
Verify that the host device is
correctly configured and
compatible with NomoLine ISA
CO2.
LEGI indicator is
dark (indicator is
off)
NomoLine ISA CO2 is not
plugged in to the host
monitor connector
Sampling line not plugged
in to NomoLine ISA CO2
completely
Internal error
Verify NomoLine ISA CO2
connection to the host monitor.
Disconnect and reconnect the
sampling line to NomoLine ISA
CO2.
The NomoLine ISA CO2 requires
service. See Service and Return
Procedure.
LEGI indicator
blinking red
Sampling line obstructed
(occlusion)
Inspect sampling line for
occlusion (blockage, kink in line).
Replace sampling line. See
Sampling Line Replacement on
page 14.
LEGI indicator
steady red
NomoLine ISA CO2
requires a power cycle
Internal error
Disconnect and reconnect the
NomoLine ISA CO2 connector at
the host device.
If persistent, the NomoLine ISA
CO2 requires service. See Service
and Return Procedure.
A hardware error is
generated during the
warm up period after being
stored at −40°C

Other manuals for Masimo NomoLine ISA CO2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Masimo NomoLine ISA CO2 and is the answer not in the manual?

Masimo NomoLine ISA CO2 Specifications

General IconGeneral
BrandMasimo
ModelNomoLine ISA CO2
CategoryMedical Equipment
LanguageEnglish

Related product manuals