NomoLine ISA CO2 Chapter 5: Specifications
www.masimo.com 26 Masimo
Gas or Vapor Gas Level CO
2
CO (Carbon Monoxide)
5
1 vol% _
1
NO (Nitrogen Monoxide)
5
0.02 vol% _
1
O
2
5
100 vol% _
2
Note 1: Negligible interference, effect included in the specification “Accuracy, all
conditions” above.
Note 2: Negligible interference with N
2
O / O
2
concentrations correctly set, effect included
in the specification “Accuracy, all conditions” above.
Note 3: Interference at indicated gas level. For example, 50 vol% Helium typically
decreases the CO
2
readings by 6%. This means that if measuring on a mixture containing 5.0
vol% CO
2
and 50 vol% Helium, the actual measured CO
2
concentration will typically be
(1-0.06) * 5.0 vol% = 4.7 vol% CO
2
.
Note 4: According to the EN ISO 80601-2-55 standard.
Note 5: In addition to the EN ISO 80601-2-55 standard.
Symbols
The following symbols may appear on the product or product labeling:
Symbol Description Symbol Description
Follow instructions for use
Consult instructions for use
Separate collection for electrical
and electronic equipment
(WEEE)
Mark of conformity to European
medical device directive
93/42/EEC
Protection against splashing of
water from all directions
ETL Intertek certification
See Declarations on Page 1 for
certifications
Caution: Federal (USA) law
restricts this device to sale by or
on the order of a physician
Defibrillation-proof. Type BF
applied part
Biohazardous Waste
Serial number