NomoLine ISA CO2 Chapter 5: Specifications
www.masimo.com 22 Masimo
* A warm up period of 10 minutes is required for the NomoLine ISA CO2 to fulfill the accuracy
specification if immediately put into use after being stored at −40°C. See Chapter 4:
Troubleshooting on page 19 for additional information.
** Not requiring partial pressure greater than 50 hPa, in accordance to IEC 60601-1-12.
*** CO
2
measurement inaccuracy shall be ± (0.3 kPa + 4 % of reading).
Physical Characteristics
Item Specification
Dimensions*
91 mm x 60 mm x 37 mm (3.6” x 2.4” x 1.4”)
Weight* Approx. 98 g (0.22 lbs)
* Excluding cable, tubing and NomoLine.
Electrical
Item Specification
Power Supply 4.5 to 5.5 VDC
< 0.9 W (normal operation)
< 1 W (peak @ 5 VDC)
Compliance
EMC Compliance
See host monitor Operator's Manual.
Safety Standards Compliance
IEC 60601-1:2005/AMD1:2012
IEC 62304:2006/AMD1:2015
EN/ISO 80601-2-55:2018