O3 Regional Oximeter Chapter 4: Operation
www.masimo.com 35 Masimo
Delta Baseline (Δbase) Settings
The following are the various setting options for Delta Baseline:
About
An informational read-only screen appears with the following definition for Delta Baseline
Settings:
Relative deficit below baseline displayed as a percentage of the baseline rSO
2
level, is the
difference between the user defined baseline rSO
2
level and current rSO
2
level. If rSO
2
level is
above the baseline, delta baseline will display as 0.
Alarms
Option Description Factory
Default
Configuration
Options
Resolution
Delta
Limit
Threshold of Δbase that triggers an
alarm.
Off
Off
-1% to -99%
-1%
Silence
Duration
Duration of the temporary suspension
of audible alarm for Δbase when the
Alarm Silence icon is pressed on
Root.
2m 30s, 1m, 2m N/A