6. Operation
6.4.2.4.3 IEEE-488 Address
The IEEE-488 Address button within the Remote application proivdes a place for the operator to set the IEEE-488 address(Figure
6.4.2.4.3). When this button is selected, a keypad will appear on the right side of the screen. The operator can input a value
between 1 and 31 to change the address of the instrument.
Figure 6.4.2.4.3 IEEE-488
6.4.2.4.4 Ethernet Settings
In the Ethernet section of the Remote application, there are buttons that correspond to Ethernet parameters. When a button is
pressed, the numeric key pad or a choice selector will be displayed on the sidebar and an appropriate number or selection can be
entered for the respective parameter. With DHCP turned off, a static IP address can be assigned. If DHCP is turned on and a lease
is obtained successfully, the ethernet parameters are greyed and inactive but show the newly assigned lease. If a DHCP server
fails to respond, DHCP will automatically turn off. Turning DHCP off allows the ethernet parameters to be edited and a static IP
address assigned.
Figure 6.4.2.4.4 Ethernet Settings
6.4.2.4.5 Serial Settings
The Serial settings, within the Remote application, can be found on page two by pressing the down arrow located at the top
of the screen (Figure 6.4.2.4.5). The Serial setup page has buttons that correspond to serial communication parameters. When
a parameter button is pressed, a choice selector will be displayed on the right side of the screen. The settings include baud rate,
data bits, stop bits, parity and echo.