65
WIKA operating instructions hydraulic high pressure controller model CPC8000-H
PN 0018708001B 11/2020 EN
EN
7. Remote Operation
:UPPer? Returns maximum control limit.
:VENT <n> Pressure at which the vent solenoid opens for an uncon-
trolled vent.
:VENT? Returns the pressure at which the vent solenoid opens.
:VRATE Sets the rate at which the controlled vent goes to
atmosphere.
:VRATE? Returns the rate at which the controlled vent goes to
atmosphere.
:SYSTem
:DETECT SLOW/FAST/CANCEL Starts or stops controller adaption routine.
Note: There is no dierence between slow and fast. Both
start the same controller adaptation process, and are only
for backwards compatibility.
:DETECT? Returns status of controller adaption routine where the
response is:
0, OK
-1, active
-2, leakage or no pressure supply
-3, pressure supply too high or wrong mode
-4, default during controller adaption
-5, calculation fault
-6, abandoned process
:TARE
:STATe ON/OFF/1/0 Tare the current reading. The same tare value is applied
to all measuring transducers.
:STATe? Returns the tare value.