Functional description
100,000 cts digital multimeters 23
2. Description of the “Keypad” keys (continued)
SURV: monitors the variations of a signal,
recording the extremes (MIN, MAX) of the
main measurement and calculating its mean
(AVG).
For each quantity stored, the multimeter records
the corresponding date and time.
When the SURV mode is entered by
Start [F1], the last MIN and MAX
measurements are erased, then
initialized with the present
measurement; to stop this mode,
press [F2] stop; [F3] to look up.
AVG is the calculated mean of all
measurements made since the activation of the
SURV mode.
The recorded data can be looked up by
pressing Look up key [F3].
In the SURV mode:
- management of the MANU or AUTO range
management cannot be selected.
- the present measurement, the MIN value,
and the MAX value are presented in the
ranges best suited to each of them.
The recorded data are accompanied by the
date and time, along with the surveillance
range.
Please update your multimeter before
starting a SURVeillance campaign
(automatic synchronization).
Reset of the MIN/MAX values by a long
press on Meas …
www.ShopAEMC.ca
Shop for AEMC products online at:
1.800.561.8187